Product Core Introduction
MAS-100 Sirius is Merck Millipore’s latest-generation high-precision active microbial air sampler, developed in full compliance with ISPE GAMP5, EU GMP Annex 1 and ISO 14698/EN 17141 regulatory standards. Built on the proven Andersen impaction principle, it delivers validated physical and biological sampling efficiency for Grade A/B pharmaceutical cleanrooms, isolators and controlled aseptic processing zones. The unit features a smart side-tilting handle design that eliminates operator hand exposure to critical sampling areas, paired with dual mass flow sensors to ensure consistent airflow accuracy within ±2.5% tolerance, fully meeting 21 CFR Part 11 data integrity requirements for regulated industries .

Key Technical Specifications
| Parameter | Specification |
|---|---|
| Official Full Designation | MAS-100 Sirius Microbiological Air Sampler |
| Manufacturer | Merck Millipore (MBV AG, Switzerland) |
| Sampling Principle | Andersen impaction method, nominal d50 cut-off value of 1.1 µm |
| Certified Airflow Rates | 100 SLPM and 200 SLPM, accuracy ±2.5% |
| Housing Material | 316L stainless steel main body, scratch-resistant glass display cover |
| Physical Dimensions | H 27.7 cm × W 14.7 cm × L 16 cm |
| Net Weight | 2.6 kg (with standard perforated lid) |
| Display Unit | Transflective glass touchscreen, sunlight-readable for cleanroom operation |
| Plate Compatibility | Tool-free adjustable holder, supports 55 mm to 90 mm agar plates |
| Communication Interfaces | Wi-Fi for remote control, USB 2.0 for data export and firmware upgrade |
| Power System | Rechargeable lithium-ion battery, 8 hours continuous sampling runtime |
| Environmental Rating | IP54, resistant to standard disinfectants including sporicidals and VHP |
| Operating Temperature | 0°C to +50°C |
| Regulatory Compliance | 21 CFR Part 11, GAMP5, PDA TR #13, EU GMP Annex 1 |
| Standard Warranty | 12 months from factory calibration date |
6 Recommended Compatible Models in the Same Series
- MAS-100 Iso NT®: Isolator-specific variant with Profinet/ProfiBus interface, designed for VHP decontamination cycle integration in restricted access barrier systems
- MAS-100 Iso MH®: Remote split control unit model, control module placed outside critical zone, sampling head deployed inside isolator with maximum 50-meter tubing extension
- MAS-100 NT®: Classic entry-grade cleanroom air sampler, 100 SLPM fixed flow rate, widely deployed in pharmaceutical fill-finish lines for routine environmental monitoring
- MAS-100 ECO®: Lightweight cost-optimized version, 1.4 kg aluminum housing, 100 SLPM flow rate with ±4% accuracy, for non-Grade A general cleanroom applications
- MAS-100 Sirius Wi-Fi Kit: Premium bundled package, including wireless remote control software, 10 sets of disposable Safe Lid and NIST-traceable airflow calibration certificate
- MAS-100 Portable Field Unit: Outdoor environmental monitoring variant, with IP65 rugged enclosure and extended 20-hour battery life, for biopharmaceutical facility ambient air survey tasks
Professional Quality Control Test Standard Operating Procedure
- Visual & Mechanical Pre-Validation
Inspect 316L stainless steel housing for surface finish compliance, verify magnetic lid closure alignment and perforated lid hole pattern integrity; confirm no scratch or deformation on sealing surfaces that could cause air leakage. - Dual Mass Flow Sensor Calibration Bench Test
Connect unit to NIST-traceable DA-100 digital anemometer, perform 10-point airflow calibration at 100 SLPM and 200 SLPM, ensure all measured values fall within ±2.5% tolerance range. - Biological Sampling Efficiency Challenge Test
Run standardized bioaerosol challenge with Bacillus atrophaeus spores, validate sampling efficiency meets ISO 14698 requirement for >50% capture rate of particles smaller than 1.62 µm. - Disinfectant Compatibility Endurance Test
Wipe all external surfaces 100 times with 75% ethanol, 0.5% peracetic acid and VHP vapor exposure, confirm no material corrosion, label fading or functional degradation after test. - Data Integrity Audit Trail Validation
Execute 50 consecutive simulated sampling cycles, verify all electronic records automatically generate timestamp, operator ID and audit trail; confirm no unauthorized modification of sampling volume or time parameters is possible. - Final Environmental Stress Release Test
Run 24-hour continuous operation at 40°C / 90% RH, perform full functional self-diagnostic; attach unique calibration serial number label, generate factory IQ/OQ documentation and release for shipment.
New Module Installation & Startup Guideline
- Unpack the unit, confirm all standard accessories (stainless steel perforated lid, power supply, USB data cable, quick start manual) are included; perform full exterior disinfection with 70% IPA before bringing into cleanroom airlock.
- Align the magnetic perforated lid with the sampling head base, release the magnetic lock until it clicks firmly; place selected 90 mm agar plate into the tool-free adjustable holder, confirm no gap between plate rim and sealing gasket.
- Connect AC power supply or insert fully charged lithium battery, power on the unit; complete initial system setup including operator ID input, cleanroom location tag assignment and real-time clock synchronization.
- Run built-in automatic leak self-test, wait for system to confirm no air leakage fault; set target sampling volume (10-2000 L selectable) and start pre-sampling 30-second purge cycle to eliminate residual background contamination.
- After sampling completion, the unit automatically stops airflow, lifts the perforated lid; export full sampling data log via USB or Wi-Fi to local LIMS system, generate 21 CFR Part 11 compliant electronic report for regulatory archiving.
Application Scenarios & Distinctive Features
- Aseptic Filling Line Grade A Zone Monitoring: Zero operator hand intrusion design eliminates secondary contamination risk during sampling, fully meeting EU GMP Annex 1 mandatory requirements for critical zone environmental monitoring.
- Isolator & RABS Integrated Monitoring: Compatible with VHP decontamination cycles, the sampling head can remain inside isolator during full sterilization process without removal, drastically reducing workflow interruption.
- Pharmaceutical Water For Injection Production Area Routine Testing: Dual flow rate 100/200 SLPM selection cuts long sampling cycle time by 50%, significantly improving environmental monitoring team daily work efficiency.
- Biotech Cleanroom Validation & Turnover Projects: Built-in Wi-Fi remote control allows engineers to operate multiple samplers simultaneously outside Grade B background zone, reducing personal protective equipment consumption and operator exposure time.
- Core Advantages: Zero-tool agar plate size adjustment, dual independent mass flow sensors for redundant safety, no laminar airflow disturbance during operation, full backward compatibility with all historical series sampling protocols.
Frequently Asked Questions
- Q: The unit keeps reporting airflow out of tolerance error during sampling?
A: Check if the perforated lid holes are blocked by residual agar or dust, perform full lid disassembly and autoclave cleaning; re-run zero-point calibration after cleaning, avoid operating the unit at altitude over 2000m without altitude compensation. - Q: Can I use third-party non-validated agar plates with this sampler?
A: Merck strongly recommends using certified 55-90 mm ICR/ICRplus settle plates; non-standard thickness plates may break the sealing gap and cause airflow deviation, leading to non-compliant sampling results. - Q: How often should I perform formal NIST-traceable airflow recalibration?
A: Per GMP requirement, annual recalibration via Merck authorized service center is mandatory; you can perform monthly field spot check using DA-100 digital anemometer to verify performance between formal calibrations. - Q: Is the sampling data directly acceptable for FDA and EU GMP audit?
A: Yes, all Sirius electronic records include immutable audit trail, timestamp and unique device serial number, fully meet 21 CFR Part 11 data integrity requirements for regulatory submission. - Q: What is the difference between Sirius and older NT generation?
A: Sirius upgrades to dual mass flow sensor, adds Wi-Fi remote control, supports 200 SLPM high flow rate and 55-90 mm universal plate compatibility; all historical sampling data format is fully backward compatible, no LIMS system modification required.