Description
HP 5517C Zeeman Dual-Frequency He-Ne Laser Head, High-Speed Heterodyne Interferometer Source
Model & Part Number
- Type: 5517C
- Manufacturer: HP (later Agilent, now Keysight)
- Product Category: Zeeman-split dual-frequency Helium‑Neon laser head
- Compatible System: 5527B laser positioning transducer, VMEbus / PC axis boards, full 5517 interferometer optics family
- Available Options: 003 (3 mm beam), 009 (9 mm beam), H03 / H05 / H07 higher power variantsKeysight

HP 5517C
Product Description
The 5517C is a high-velocity dual-frequency He‑Ne laser head in the classic 5517 series, upgraded from the 5517B for faster motion axes. It adopts Zeeman splitting to generate a beat frequency of 2.4–3.0 MHz, supporting linear measurement velocity up to 711 mm/s, ideal for high-speed wafer stages and FPD motion platforms. The internal reference photodiode continuously locks the optical frequency to maintain ultra-stable vacuum wavelength 632.99137 nm. Identical mechanical footprint and mounting interface as the 5517B for drop-in replacement when higher tracking speed is required. It is end-of-life legacy metrology hardware, and aftermarket demand comes from semiconductor fabs, precision motion OEMs and calibration labs for spare swap-out and system life extensionKeysight.
Technical Specifications
| Item | Specification |
|---|---|
| Laser Type | CW He‑Ne, Zeeman dual-frequency |
| Nominal Vacuum Wavelength | 632.99137 nm |
| Beat Reference Frequency | 2.4–3.0 MHz |
| Max Linear Velocity | 711 mm/s |
| Standard Beam Diameter | 6 mm; Opt 003:3 mm; Opt 009:9 mm |
| Beam Height | 79.5 ±1.0 mm |
| Optical Output Power | 180 μW min ~ 1 mW max |
| Wavelength Accuracy | ±0.1 ppm (3σ lifetime); ±0.02 ppm with MIL‑STD 45662 calibration |
| Wavelength Stability | ±0.002 ppm / 1 hour typical; ±0.02 ppm lifetime |
| Power Supply | +15 VDC ±0.3 V, max 3.0 A; −15 VDC ±0.3 V, max 0.3 A |
| Heat Dissipation | 23 W steady operation; 35 W during warm-up |
| Warm-up Time | 90 minutes |
| Weight | 3.4 kg (7.5 lb) |
| Operating Temperature | 0°C ~ +40°C; 17–23°C recommended |
| Storage Temperature | -20°C ~ +70°C |
| Relative Humidity | 5% ~ 85%, non-condensing |
| Laser Safety Class | Class II (21 CFR 1040.10 / 1040.11) |
| MTBF | > 50,000 hours typical |
| Compliance | CE |
Application Fields
- Semiconductor lithography and high-speed wafer positioning stages
- Flat panel display inspection and exposure motion platforms
- High-precision CNC machine tool calibration for fast moving axes
- Coordinate Measuring Machines (CMM) with high traverse speed
- Aerospace ultra-precision positioning and metrology
- Precision grinding and high-speed linear motor stages
- Metrology institutes requiring high-bandwidth interferometer measurement
- Legacy 5527 interferometer system maintenance & spare replacement
Brand Profile (HP / Agilent / Keysight)
HP invented and commercialized Zeeman dual-frequency laser interferometry for industrial position feedback. The 5517 product family became the global standard for heterodyne metrology. After HP spun off its test & measurement division into Agilent and later Keysight, the 5517 series remained the benchmark. The 5517C fills the performance gap between the 5517B and top-tier 5517D, targeting high-speed precision motion widely used in semiconductor and display industries. Modern laser metrology platforms are available, but massive installed fab equipment still relies on 5517C spare inventory.
Related News & Industry Events
The 5517C is officially obsolete by Keysight. Factory new stock is exhausted and lead times are extremely long. Semiconductor and FPD factories keep calibrated spare 5517C units on hand, because laser tube decay can halt critical exposure and inspection tools. Many users evaluate system upgrade during tool overhaul, while retaining tested 5517C units to avoid production interruption. Fully tested, wavelength-verified 5517C laser heads remain high-value aftermarket metrology spares.
Usage Scenarios
- Stabilized light source for high-speed heterodyne laser interferometer systems
- Position feedback for semiconductor wafer scanner and inspection stages
- Linear, straightness, pitch/yaw and squareness calibration of fast-travel machine axes
- Direct drop-in upgrade / replacement for 5517B when higher tracking velocity is needed
- Multi-axis volumetric error mapping for high-speed CNC and CMM systems
- Reference wavelength standard for calibration laboratories
- Optical source matched with 5527B transducer and VME/PC axis interface cards
- Spare laser head for legacy FPD and semiconductor fab motion metrology racks
Functional Advantages
- Higher 2.4–3.0 MHz beat frequency, supports up to 711 mm/s tracking speed (faster than 5517B)
- Zeeman heterodyne architecture, excellent noise rejection and nanometer measurement resolution
- Built-in reference detector for continuous wavelength locking
- Same mechanical footprint as 5517B, simple retrofit without bracket modification
- Multiple beam diameter options (3/6/9 mm) to match different interferometer optics
- Class II low-power laser, safe for normal lab and fab operation
- Fully compatible with the complete Keysight 5517 series optics and receiver ecosystem
- Proven long-term stability for 24/7 fab and metrology operation
- Calibration certificates available for ISO quality and traceability requirements
Recommended Related Products (16 pcs)
- Keysight 5517A Laser Head
- Keysight 5517B Laser Head
- Keysight 5517D Laser Head
- Keysight 5517CL Forced Air Cooling Laser Head
- Keysight 5517DL Forced Air Cooling Laser Head
- Keysight 5527B Laser Positioning Transducer
- Keysight 10885A PC Axis Board
- Keysight 10895A VMEbus Axis Board
- Keysight 10706A Linear Interferometer
- Keysight 10707A Plane Mirror Interferometer
- Keysight 10715A Retroreflector
- Keysight 10780C Optical Receiver
- Keysight 10719A Beam Bender
- Keysight 10721A Straightness Optics
- Keysight 5517 Series Calibration Service
- Keysight 10735A Fiber Optic Launcher
Installation Instruction
- Mount the laser head on a vibration-isolated optical base or tool bracket.
- Disconnect +15 V / −15 V DC power before wiring.
- Connect dual DC power supply strictly following polarity and current rating.
- Align the outgoing laser beam to interferometer optics; avoid direct reflected light back into the laser aperture.
- Keep the beam path clean and minimize air turbulence.
- Apply power and allow full 90-minute warm-up before metrology measurement.
- Confirm lock status and stable beat frequency within 2.4–3.0 MHz.
- Align interferometer optics and optical receiver to obtain strong heterodyne signal.
- Connect to axis board / 5527B transducer and verify position reading.
- Run high-speed velocity test, confirm no signal dropout up to 711 mm/s.
- Archive wavelength data and calibration certificate for quality records.
- Apply laser safety label and beam guarding according to Class II laser requirements.
Maintenance Guide
- Daily Check: Inspect laser lock indicator, beam brightness and housing temperature; avoid direct eye exposure to laser beam.
- Periodic Maintenance: Clean output window only with professional optical lens tissue; do not touch the coated surface.
- Routine Test: Perform annual calibration to verify wavelength and beat frequency; validate measurement accuracy.
- Troubleshooting: Loss of lock is usually caused by aging He‑Ne tube, vibration, unstable ±15V supply or optical feedback. Low optical power indicates end-of-life laser tube.
- Spare Handling: Store spare 5517C in dry, low-vibration environment; protect optical window from dust and scratches.
- Repair Notice: The He‑Ne laser tube is hermetically sealed; no on-site repair possible. Return unit to specialized metrology vendor for tube replacement and recalibration.
- Lifecycle Note: 5517C is obsolete metrology hardware. Evaluate modern laser interferometer upgrade while maintaining calibrated spare inventory for fab and calibration continuity.