HP 5517B Dual-Frequency He-Ne Laser Head for Laser Interferometer Metrology System

  1. Keysight 5517A Laser Head
  2. Keysight 5517C Laser Head
  3. Keysight 5517D Laser Head
  4. Keysight 5517BL Forced Air Cooling Laser Head
  5. Keysight 5527B Laser Positioning Transducer
  6. Keysight 10885A PC Axis Board
  7. Keysight 10895A VMEbus Axis Board
  8. Keysight 10706A Linear Interferometer
  9. Keysight 10707A Plane Mirror Interferometer
  10. Keysight 10715A Retroreflector
  11. Keysight 10780C Optical Receiver
  12. Keysight 10719A Beam Bender
  13. Keysight 10721A Straightness Optics
  14. Keysight 5517 Series Calibration Service
  15. Keysight 10735A Fiber Optic Launcher
  16. Keysight 5517 Series Power Supply Module
Category: SKU: HP 5517B Brand:

Description

HP 5517B Dual-Frequency He-Ne Laser Head for Laser Interferometer Metrology System

 

Model & Part Number

  • Type: 5517B
  • Manufacturer: HP (Hewlett-Packard), later Agilent, now Keysight
  • Product Category: Zeeman-split dual-frequency Helium-Neon laser head
  • Compatible System: Keysight 5527B laser positioning transducer, VMEbus & PC-based interferometer axis boards, 10885A / 10895A axis modules
  • Variant Option: H10 (2.2–2.4 MHz beat frequency, higher optical power)

HP 5517B

Product Description

The 5517B is a Class II dual-frequency He-Ne laser source designed for high-precision laser interferometer displacement measurement. It outputs a stabilized 632.99137 nm vacuum wavelength with Zeeman splitting to produce a beat frequency of 1.9–2.4 MHz, supporting linear measurement velocity up to 500 mm/s. The integrated laser tube, permanent magnet assembly and internal reference photodetector continuously lock optical frequency to maintain ultra-low wavelength drift. It is widely used as the light source for machine tool calibration, CMM and semiconductor stage positioning. This laser head is obsolete / end-of-life, and aftermarket demand is driven by metrology lab, precision machine and fab spare replacement and system life extension.

 

Technical Specifications

Item Specification
Laser Type CW He-Ne, Zeeman dual-frequency
Nominal Vacuum Wavelength 632.99137 nm
Beat Reference Frequency 1.9–2.4 MHz (Option H10: 2.2–2.4 MHz)
Max Linear Velocity 500 mm/s
Beam Diameter 6 mm
Beam Height 79.5 ±1.0 mm
Optical Output Power 120 μW min ~ 1 mW max
Wavelength Accuracy ±0.1 ppm (3σ lifetime); ±0.02 ppm with MIL-STD 45662 calibration
Wavelength Stability ±0.002 ppm / 1 hour typical; ±0.02 ppm lifetime
Power Supply +15 VDC ±0.3 V, max 3.0 A; −15 VDC ±0.3 V, max 0.3 A
Heat Dissipation 23 W steady operation; 35 W during warm-up
Warm-up Time 90 minutes (standard version)
Weight 3.4 kg (7.5 lb)
Operating Temperature 0°C ~ +40°C; 17–23°C recommended for optimal linearity
Storage Temperature -20°C ~ +70°C
Relative Humidity 5% ~ 85%, non-condensing
Laser Safety Class Class II (21 CFR 1040.10 / 1040.11)
MTBF > 50,000 hours typical
Compliance CE

 

Application Fields

  • CNC machine tool linear and geometric calibration
  • Coordinate Measuring Machine (CMM) position feedback
  • Semiconductor lithography and wafer stage positioning
  • Precision metrology laboratories and calibration institutes
  • Flat panel display inspection and motion stages
  • Aerospace precision component measurement
  • Ultra-precision grinding and turning equipment
  • Legacy laser interferometer system maintenance & spare replacement

 

Brand Profile (HP / Agilent / Keysight)

HP pioneered dual-frequency laser interferometry for industrial metrology. The 5517 series became the industry standard for optical displacement measurement. After HP’s test & measurement division spun off as Agilent, and later Keysight, the 5517 family remained the benchmark for stable He-Ne laser heads. The 5517A/B/C/D lineup covers different beat frequencies and slew rates for low to high-speed motion axes. Modern metrology platforms have adopted newer laser sources, while a huge installed base of calibrated machines still depends on 5517B spares.

 

Related News & Industry Events

5517B is officially obsolete by Keysight. Original factory new stock is exhausted, and replacement lead times are extremely long. Metrology labs and high-end machine shops maintain tested spare 5517B units to avoid downtime of calibration equipment. Laser tube decay is the main failure mode for aged heads. Many users evaluate upgrading to newer laser interferometer systems during overhaul, while keeping calibrated surplus 5517B units to sustain existing metrology workflows. Fully tested and wavelength-verified 5517B remain high-value aftermarket items.

 

Usage Scenarios

  1. Primary stabilized light source for dual-frequency laser interferometer systems
  2. Linear positioning, pitch, yaw, straightness and squareness machine calibration
  3. Closed-loop position reference for precision motion stages
  4. Direct replacement for failed original 5517B laser head in metrology racks
  5. Multi-axis interferometer setup for large machine tool volumetric error mapping
  6. Reference standard for calibration laboratory displacement verification
  7. Optical source for 5527B laser transducer and VMEbus axis interface cards
  8. Spare laser head for semiconductor and display fab legacy metrology stations

 

Functional Advantages

  • Zeeman-split dual-frequency design, robust for heterodyne interferometer measurement
  • Excellent short-term and long-term wavelength stability for nanometer-level metrology
  • Built-in reference detector for automatic frequency locking
  • Class II low-power laser, safe for normal industrial handling
  • Standard 6 mm beam, compatible with full set of Keysight interferometer optics
  • Mature matched ecosystem with axis boards, receivers and optics from the 5527 system
  • High MTBF, proven decades-long service life in 24/7 precision metrology
  • Fixed beam height and footprint, drop-in mechanical replacement for original 5517B
  • Traceable wavelength calibration available for quality system compliance

 

Recommended Related Products (16 pcs)

  1. Keysight 5517A Laser Head
  2. Keysight 5517C Laser Head
  3. Keysight 5517D Laser Head
  4. Keysight 5517BL Forced Air Cooling Laser Head
  5. Keysight 5527B Laser Positioning Transducer
  6. Keysight 10885A PC Axis Board
  7. Keysight 10895A VMEbus Axis Board
  8. Keysight 10706A Linear Interferometer
  9. Keysight 10707A Plane Mirror Interferometer
  10. Keysight 10715A Retroreflector
  11. Keysight 10780C Optical Receiver
  12. Keysight 10719A Beam Bender
  13. Keysight 10721A Straightness Optics
  14. Keysight 5517 Series Calibration Service
  15. Keysight 10735A Fiber Optic Launcher
  16. Keysight 5517 Series Power Supply Module

 

Installation Instruction

  1. Secure the laser head on a vibration-isolated optical base or machine bracket.
  2. Disconnect +15 V / −15 V DC power before wiring.
  3. Connect dual DC power supply strictly following polarity and current rating.
  4. Align the outgoing laser beam to the interferometer optics path; avoid specular reflection back into the laser aperture.
  5. Ensure the beam path is clean and free of obstructions and air turbulence.
  6. Apply power and allow full 90-minute warm-up before metrology work.
  7. Monitor lock status; confirm beat frequency is stable within 1.9–2.4 MHz range.
  8. Align interferometer optics and receiver to obtain strong heterodyne signal.
  9. Connect to axis board / 5527B transducer and verify position reading.
  10. Perform velocity test to confirm no signal dropout up to 500 mm/s.
  11. Record wavelength data and calibration certificate for quality records.
  12. Add safety label and beam path guarding per Class II laser requirements.

 

Maintenance Guide

  • Daily Check: Check laser status indicator, beam brightness, no abnormal heating. Avoid looking directly into laser beam.
  • Periodic Maintenance: Keep output aperture clean with professional optical lens tissue; do not touch the window surface.
  • Routine Test: Recheck wavelength and beat frequency during annual calibration; validate measurement accuracy against reference standard.
  • Troubleshooting: Loss of lock usually indicates aging laser tube, excessive vibration, unstable ±15V supply or optical feedback. Low optical power is typical end-of-life symptom of He-Ne tube.
  • Spare Handling: Store spare 5517B in dry, low-vibration environment; protect optical window from dust and scratches.
  • Repair Notice: He-Ne laser tubes are sealed; on-site repair is not possible. Return unit to specialized metrology vendor for laser tube replacement and recalibration.
  • Lifecycle Note: 5517B is obsolete metrology hardware. Evaluate modern laser interferometer upgrade while maintaining calibrated spare inventory for continuity of calibration services.